Every one an expert
eVol® xr digital analytical syringe is the coupling of two precision devices: a digitally controlled electronic drive and an XCHANGE® enabled analytical syringe.
eVol xr syringes are easily and quickly changed allowing them to be dedicated to individual liquids or methods to prevent possible cross-contamination of reagents.
Features and benefits
- Syringes compatible for use with eVol xr.
- Suitable for use with volatile samples.
- Variable speed of operation.
- Easily calibrated and calibration factors saved for each syringe, enabling laboratories to comply with stringent global laboratory standards (e.g. GLP, GMP, FDA).
- Programmable and able to store a laboratory workflow (up to 98 steps).
- Password protection options enabling standardization of work processes.
- Reduces the need for serial dilutions.
Recommended applications
- Preparation and addition of standards.
- Precise dispensing of aqueous and non-aqueous liquids.
- TLC spotting.
- Micro titrations.
- GC and LC instrument injections.
- NMR sample preparation.
- MEPS®
Product specifications
- 5 µL eVol xr syringes available with 50 mm and 115 mm long needles.
- 50 µL eVol xr syringes available with 50 mm, 115 mm, and 180 mm long needles.
- 100 µL eVol xr syringes available with a 50 mm long needle.
- 500 µL eVol xr syringes available with 50 mm, 70 mm, 115 mm, and 180 mm long needles.
- 1 mL eVol xr syringes available with 50 mm, and 70 mm long needles.
- 50 µL, 100 µL and 500 µL removable needle eVol xr MEPS syringes available to fit MEPS BINs.
Volume range of eVol xr syringes
Accuracy of eVol xr syringes
Syringe volume | 1 mL | 500 µL | 100 µL | 50 µL | 5 µL |
---|---|---|---|---|---|
At 100% (uncalibrated)* | ±0.7% | ±0.7% | ±0.7% | ±1.0% | ±1.0% |
At 100% (calibrated)** | ±0.2% | ±0.2% | ±0.2% | ±0.2% | ±0.2% |
At 10% (calibrated)*** | ±0.5% | ±0.5% | ±0.5% | ±0.5% | ± 1.0% |
* Calibration factor of 1.0000 applied to handle.
** Syringe and handle calibrated at full volume.
*** Syringe and handle calibrated at 10% of capacity. Accuracy based on the 10% dispensed volume closest to zero position.
Note: For best results when performing repeat dispense (one aspiration followed by multiple dispense steps) use a calibration factor of 1.0000.
Precision of eVol xr syringes
Syringe volume | 1 mL | 500 µL | 100 µL | 50 µL | 5 µL |
---|---|---|---|---|---|
At 100% | ±0.3% | ±0.3% | ±0.4% | ±0.4% | ±0.5% |
At 10% | ±0.6% | ±0.6% | ±0.7% | ±0.8% | ±1.0% |
Specifications are subject to change without notification.
More
Please refer to the part number list for more items.
2 products
Brochure - Syringes for the laboratory
Collections: Agilent Technologies, All, Analytical, Automation, Automation consumables, Autosampler syringes, Brochures, CTC Analytics, CTC RTC, Dispensing syringes, eVol syringes, L-MARK syringes, Literature - All, Literature - Analytical, Literature - Automation, Manual syringes, PerkinElmer, Plunger-in-needle syringes, Select by instrument - Analytical, SGE pump dispenser syringes, Shimadzu, Shimadzu GC 2010, Shimadzu GC 2010/17A, Support items, Syringes, Thermo Scientific, Thermo Scientific AS2000, Thermo Scientific AS3000, Thermo Scientific TRACE 1300 GC, Thermo Scientific TRACE 1300 Series GC, Thermo Scientific TriPlus, Thermo Scientific TriPlus RSH
Part Number List - MEPS syringes
Collections: Agilent Technologies, All, Analytical, CTC Analytics, eVol syringes, Literature - All, Literature - Analytical, MEPS, Part Number Lists, Select by instrument - Analytical, Shimadzu, Support items, Syringes, Thermo Scientific